# Operators Guide

Last Revised: June 2024
Author: Pamela Schleissner
**IN CASE OF EMERGENCY PRESS F12**
--> Enables attenuator 11: att(1e<sup>-16</sup>)

## Novice User

1. Starting and using the DAQ (data acquisition system):

a. **restartdaq** -- Launches daq, **restartdaq -w** launches DAQ
with the windows in their previous position

b. **startami** -- restarts ami, use if daq is working but ami is
frozen

- Load the window configuration

c. It is good practice to let the DAQ run freely to get updated
detector and AMI plots in between scans without recording
(**daq.disconnect()** and then "begin run" from dropdown menu in
DAQ window, with unchecked "Record Run"). Pulse picker needs to
be in **'open'** or **'flip flop'** mode for observations

2. MFX Home page:

a. **mfxhome** -- Launches MFX home

b. Checking for faults you may have to clear if the beam goes down:
LCLS → LFE Home → General tab → PMPS Diagnostics → clear faults

3. MFX Python:

a. From a seperate terminal **mfx3** starts a python session

b. If python freezes: **exit()** then **mfx3** to restart

4. Open **camViewer** and **eloggrabber** from a terminal

5. X-ray attenuation: **att(1e-16)** max. **att(1)** min attenuation →
100% transmission

6. Moving the Rayonix detector:

![](./media/ops_guide_rayonix2.png)
![](./media/ops_guide_rayonix.png)

a. **Use small step sizes!** If you move the detector in one step, you
will crash the detector into the spectrometer, because it overshoots
in the end!

b. Pull out the flight tube out (slowly) before moving the detector in

c. Remember to put the flight tube in against the kapton before data
collection

7. Pedestals:

a. For epix100 and epix10k2M:

i. From a terminal: **takepeds** (will take control of daq,
need to disconnect DAQ if not done before), confirm X-rays
are blocked (**stopper S4.5 is inserted**) and it will start
taking pedestals

ii. Once takepeds is done, run **makepeds** following the prompt
(eg. **makepeds -u \<userid\> -q milano -r 28** )

b. For the Rayonix detector:

i. From a terminal:\
\>\> **ssh -X hsuser@con-ics-mfx-rayonix**\
\>\> **killall procServ**\
\>\> **capxure**\
\
A new window will appear. Change the "single trigger" option
to "none". Change "exposure" to "dark". Click "collect new
background". Change "exposure" to "normal". Click "collect
new background". Change "single trigger" back to frame. Exit
this window, file → quit\
\
In the previous terminal: \>\>
**./startDaqInterface_newcradyl**

c. *After collecting and deploying a pedestal, allocate to the
DAQ.* In the DAQ window: Shutdown → Allocate → Begin Running

8. Running Scans:

a. autorun(sample = '<SAMPLE TAG>', run_length = 300,
record=True, runs = 5, inspire=False, daq_delay=5,
picker='flip')

i. <Sample tag>: user defined

ii. run_length=300: in seconds

iii. record=True: records data in DAQ

iv. runs = 5: number of runs

v. inspire=False: set to True if you need some inspiration to
keep hope alive

vi. daq_delay=5: delay time between runs. Increase is the DAQ is
being slow

vii. picker='flip': If 'open' pulse picker opens before the run
starts. If 'flip' it flip-flops before run starts

## Troubleshooting 😵‍

Don't see X-rays?

1. Did you remove the stopper 4.5?

2. Is the pulse picker in "flip-flop"?

3. Are you attenuated?

4. Check "are we ready?" In a terminal type **awr mfx**

Are the plots frozen (AMI)?

1. Try shutting down the daq: In DAQ window **Shutdown****Allocate**
**Begin Running**

2. Try restarting AMI: in a terminal that is not running python
**startami**

3. Try restarting the DAQ: in a terminal window that is not running
python **restartdaq -w**

4. Call beamline technical support

Is the DAQ frozen?

1. Try restarting the DAQ: in a terminal window that is not running
python **restartdaq -w**

2. Call beamline technical support

Did the pulse picker lose its position?

1. On MFX home, open the pulse picker tab

2. Click "home motor". It will go in "home motor mode"

3. Click "home motor" (again). The motor will begin moving

4. Wait until it's done moving

5. Then hit "Reset Mode"

![](./media/ops_guide_pp.png)
![](./media/ops_guide_pp2.png)

## Experienced Users

1. Plotting AMI

2. Running CCTBX

a. Edit cctbx.xfel file by changing the experiments name (e.g.
xcsl######)

b. In a terminal that is running python: **source
\<path\>/cctbx.xfel**

c. A new window will appear with the cctbx.XFEL login. Click OK.

d. Change the trial number in the options page. Click on "Auto plot
last five runs"

e. Click "Hide Options" to see plots. Optional "Large text" for big
screens.

3. Running OM

Binary file added docs/media/ops_guide_pp.png
Loading
Sorry, something went wrong. Reload?
Sorry, we cannot display this file.
Sorry, this file is invalid so it cannot be displayed.
Binary file added docs/media/ops_guide_pp2.png
Loading
Sorry, something went wrong. Reload?
Sorry, we cannot display this file.
Sorry, this file is invalid so it cannot be displayed.
Binary file added docs/media/ops_guide_rayonix.png
Loading
Sorry, something went wrong. Reload?
Sorry, we cannot display this file.
Sorry, this file is invalid so it cannot be displayed.
Binary file added docs/media/ops_guide_rayonix2.png
Loading
Sorry, something went wrong. Reload?
Sorry, we cannot display this file.
Sorry, this file is invalid so it cannot be displayed.
2 changes: 1 addition & 1 deletion mkdocs.yml
Original file line number Diff line number Diff line change
Expand Up @@ -6,7 +6,7 @@ nav:
- Home: index.md
- Quick Start: Quick_Start.md
- Operating Procedures:
- General.md
- Operators_Guide.md
- DoT.md
- Source:
- MFX:
Expand Down