## Calibration Script

There is a very small script to determine the Factor and the Formula:
Put a voltmeter to measure the voltage between GND and Po to measure voltage.
You need the measured voltage while producing a shortcut between inner pole and outer area of the BNC of the probeboard.
You need the measured voltage when measuring the buffer pH 4.01
- Put a voltmeter to measure the voltage between GND and Po to measure voltage.
- You need the measured voltage while producing a shortcut between inner pole and outer area of the BNC of the probeboard.
- You need the measured voltage while measuring the buffer pH 4.01.

`cd /home/pi/craftbeerpi3/modules/plugins/PHMeasureADS1115`

`python calibration.py`

Follow instructions.
- Follow instructions.

## The formula

Expand Down Expand Up @@ -164,7 +164,7 @@ There is a very small script to determine the Factor and the formula:

Use this sensor as any other sensor in Craftbeerpi 3.
The Digit and Voltage values can help to calibrate. They are not needed for pH measurement.
The main calibaration is already described above.
The main calibaration is already described above and more precise at he end of this file.

Keep in mind that it takes several minutes to get the right pH value.