@@ -4,7 +4,7 @@ Now that we have assembled the OpenSPIM microscope it is time to start using it.
44

55
# Sample Preparation
66

7-
[Pulling Drosophila embryos in agarose inside a glass capillary (*with some priceless random audio*)](https://youtu.be/G6P52G06oPw)
7+
{% include video id="G6P52G06oPw" width="400" height="243" caption="Pulling Drosophila embryos in agarose inside a glass capillary (*with some priceless random audio*)" %}
88

99
The basic principles of sample preparation for SPIM differs from traditional microscopy technologies. There are no glass slides or coverslips. Instead, the sample needs to be suspended in a water filled chamber in front of the lens so that it can be rotated. This is usually achieved by embedding the sample in a low melting point (LMP) agarose inside a glass capillary equipped with a plunger. For imaging, the agarose column with the specimen is pushed out of the capillary using the plunger to hang in front of the lens.
1010

@@ -14,7 +14,7 @@ Follow the link to check our resources for [**sample preparation**](Sample_Prepa
1414

1515
# Calibration
1616

17-
[OpenSPIM rendering with laser on](images/1I_1D_OpenSPIM.png "OpenSPIM rendering with laser on")
17+
{% include image src="1I_1D_OpenSPIM.png" width="70%" caption="OpenSPIM rendering with laser on" %}
1818

1919
Before using the OpenSPIM the light sheet needs to be [**aligned**](Light-sheet_Calibration).
2020
This means that the light sheet needs to be shaped by the optics of the system to be parallel to the imaging plane of the camera, perpendicular to the detection axis, as thin as possible, uniform across the field of view and, most importantly, in focus with the detection objective. Since the procedure is rather involved, we provide a series of detailed videos that illustrate the process. Innovations are welcome.
@@ -29,7 +29,7 @@ It is also possible to measure the [light-sheet thickness](Light_sheet_character
2929

3030
# Software
3131

32-
[Screenshot of OpenSPIM stage control](images/Stagecontrols.png "Screenshot of OpenSPIM stage control")
32+
{% include image src="Stagecontrols.png" width="70%" caption="Screenshot of OpenSPIM stage control" %}
3333

3434
OpenSPIM relies on [µManager](https://micro-manager.org/wiki/) run under [Fiji](https://fiji.sc). Make sure you [**download and install**](Downloads) all the required components.
3535

@@ -40,6 +40,7 @@ Next, please follow the instructions to [**configure the hardware**](Downloads#I
4040
Make yourself familiar with the operation of the OpenSPIM [**stage control**](OpenSPIM_stage_control).
4141

4242
Now you are ready to [**acquire**](Acquisition) images.
43+
4344
Please follow these step-by-step tutorials:
4445

4546
- [Acquiring a Single Image](Acquisition#Acquiring_a_Single_Image)

‎Pre-requisites.md

+1-1
Original file line numberDiff line numberDiff line change
@@ -29,7 +29,7 @@ The processing described in this tutorial is performed on the following hardware
2929
`Network : Intel X520-T2 Dual Port `**`10GbE`` ``Network``
3030
``Interface`` ``Card`**
3131

32-
This is clearly a monster and it is not cheap (roughly 5000 Euros). Fortunately OpenSPIM data are not as humongous as SPIM data can be (for example from the commercial [Zeiss Lightsheet Z.1](https://www.zeiss.com/microscopy/int/products/imaging-systems/light-sheet-microscope-for-lsfm-imaging-of-live-and-cleared-samples-lightsheet-7.html), primarily due to limited frame rate achievable by the OpenSPIM. The configuration we recommend for processing OpenSPIM data should have:
32+
This is clearly a monster and it is not cheap (roughly 5000 Euros). Fortunately OpenSPIM data are not as humongous as SPIM data can be (for example from the commercial [Zeiss Lightsheet Z.1](http://www.zeiss.com/microscopy/en_de/products/imaging-systems/lightsheet-z-1.html), primarily due to limited frame rate achievable by the OpenSPIM. The configuration we recommend for processing OpenSPIM data should have:
3333

3434
`4 processors`
3535
`48 GB of RAM`

‎Requirements.md

+1-2
Original file line numberDiff line numberDiff line change
@@ -9,7 +9,6 @@ Hardware:
99

1010
Software:
1111

12-
- Windows (both XP and 7 are fine; needed for the Picard Stage and the
13-
Hammamatsu Orca camera)
12+
- Windows (both XP and 7 are fine; needed for the Picard Stage and the Hammamatsu Orca camera)
1413
- Fiji
1514
- µManager

‎Rotational_axis_calibration.md

+7-54
Original file line numberDiff line numberDiff line change
@@ -1,68 +1,21 @@
11
---
22
---
3-
Setting your twister angle to 0 before starting will simplify the second
4-
part below. Find a convenient bead in the live window and click
5-
Calibrate on the OpenSPIM plugin.
3+
Setting your twister angle to 0 before starting will simplify the second part below. Find a convenient bead in the live window and click Calibrate on the OpenSPIM plugin.
64

75
## Pixel Size
86

9-
The top pane helps determine pixel size, and has to be filled out before
10-
you can use the bottom pane. Using the bead you have, drag a box with
11-
one corner in the approximate center. Move the stage a known amount in X
12-
and Y -- using the arrow keys is advised as each press is 10 motor
13-
steps. Align the opposite corner of the ROI rectangle on the bead's new
14-
position, then click the Pick ROI button. It should change to show the
15-
dimensions of the selection. In the width and height fields, enter the
16-
distance the stage has traveled in motor steps (ignore the um tag right
17-
now). Click Recalculate; a value should appear next to "um per pixel:"
18-
in the Calculated Results area.
7+
The top pane helps determine pixel size, and has to be filled out before you can use the bottom pane. Using the bead you have, drag a box with one corner in the approximate center. Move the stage a known amount in X and Y -- using the arrow keys is advised as each press is 10 motor steps. Align the opposite corner of the ROI rectangle on the bead's new position, then click the Pick ROI button. It should change to show the dimensions of the selection. In the width and height fields, enter the distance the stage has traveled in motor steps (ignore the um tag right now). Click Recalculate; a value should appear next to "um per pixel:" in the Calculated Results area.
198

209
## Rotational Axis
2110

22-
The bottom pane is the 'fun' part. Enter the current twister angle **in
23-
motor steps** (double-gradians; 0 is still 0, otherwise multiply the
24-
angle in degrees by 100/180 (5/9).). Enter a reasonable interval for
25-
delta-theta -- larger values should be more accurate, but will be much
26-
more involved to calibrate for. Make sure you're at theta-zero. Enclose
27-
your convenient bead in a rectangular ROI, making certain to leave some
28-
room around the edges (this is so the gaussian fitter knows what value
29-
is 'background'). Make sure the bead is focused by scrolling in and out
30-
a step or two to find the most intense slice. Then click the first Pick
31-
Bead button. It should replace its text with the location of the bead.
32-
(You'll probably need to stretch the calibrator window quite a bit.)
11+
The bottom pane is the 'fun' part. Enter the current twister angle **in motor steps** (double-gradians; 0 is still 0, otherwise multiply the angle in degrees by 100/180 (5/9).). Enter a reasonable interval for delta-theta -- larger values should be more accurate, but will be much more involved to calibrate for. Make sure you're at theta-zero. Enclose your convenient bead in a rectangular ROI, making certain to leave some room around the edges (this is so the gaussian fitter knows what value is 'background'). Make sure the bead is focused by scrolling in and out a step or two to find the most intense slice. Then click the first Pick Bead button. It should replace its text with the location of the bead. (You'll probably need to stretch the calibrator window quite a bit.)
3312

34-
Now that you have the first location of the bead, you have to follow it
35-
through two rotations. The easiest way to do this is to use the motor
36-
controls on the SPIM plugin; increment the angle slider by one, refocus
37-
on the bead, repeat until you've changed angles by delta-theta decided
38-
above. (You may wish to click Goto theta-one to make sure you are at the
39-
correct angle.) Repeat the process of picking the bead, clicking on the
40-
second Pick Bead button instead. Repeat all of this for the third and
41-
final bead location -- you've now enough information to determine the
42-
rotational axis.
13+
Now that you have the first location of the bead, you have to follow it through two rotations. The easiest way to do this is to use the motor controls on the SPIM plugin; increment the angle slider by one, refocus on the bead, repeat until you've changed angles by delta-theta decided above. (You may wish to click Goto theta-one to make sure you are at the correct angle.) Repeat the process of picking the bead, clicking on the second Pick Bead button instead. Repeat all of this for the third and final bead location -- you've now enough information to determine the rotational axis.
4314

4415
## Final Steps
4516

46-
Once you have a pixel size and all three bead locations, you may need to
47-
click Recalculate. You should have 'reasonable' values for each
48-
calculated result: um per pixel is probably in the range of 0.4348
49-
(varies depending on stage motors). A good rotational axis has minimal
50-
values for X and Z, and a value very close to 1 (or -1) for Y -- it's
51-
very near vertical. Rotational axis origin is difficult to prescribe a
52-
'reasonable' value; the best I can suggest is that the X component
53-
shouldn't be far from your sample column.
17+
Once you have a pixel size and all three bead locations, you may need to click Recalculate. You should have 'reasonable' values for each calculated result: um per pixel is probably in the range of 0.4348 (varies depending on stage motors). A good rotational axis has minimal values for X and Z, and a value very close to 1 (or -1) for Y -- it's very near vertical. Rotational axis origin is difficult to prescribe a 'reasonable' value; the best I can suggest is that the X component shouldn't be far from your sample column.
5418

55-
After all of this, you may wish to click Save & Apply. The values will
56-
all be recorded; they'll be loaded the next time you click Calibrate --
57-
the pixel size information should be fixed, but if you've touched the
58-
sample or holder at all, you should at minimum redetermine the
59-
rotational axis.
19+
After all of this, you may wish to click Save & Apply. The values will all be recorded; they'll be loaded the next time you click Calibrate -- the pixel size information should be fixed, but if you've touched the sample or holder at all, you should at minimum redetermine the rotational axis.
6020

61-
Now that you have these values, you should be able to get a multi-view
62-
rotation by holding Alt and moving the mouse back and forth over the
63-
live window. Be patient with it -- the positional motors aren't as quick
64-
as the twister, so they sometimes lag behind a bit. With a carefully
65-
calibrated system, the X and Y should remain fairly constant on the
66-
center of the image, though the system often becomes unfocused with
67-
longer rotations. If the direction of rotation seems wrong, you can try
68-
the Reverse Axis button on the calibrator.
21+
Now that you have these values, you should be able to get a multi-view rotation by holding Alt and moving the mouse back and forth over the live window. Be patient with it -- the positional motors aren't as quick as the twister, so they sometimes lag behind a bit. With a carefully calibrated system, the X and Y should remain fairly constant on the center of the image, though the system often becomes unfocused with longer rotations. If the direction of rotation seems wrong, you can try the Reverse Axis button on the calibrator.

0 commit comments